Yerevan State University, Department of Radiophysics ; Department of Physics, Sogang University, Seoul 121-742, Korea ; Department of Physics, Sam Houston State University, Huntsville, Texas 77341, USA
We present a microwave sensor based on the modified first order Hilbert curve designed for detecting glucose concentration in aqueous solutions by using a real-time microwave near-field electromagnetic interaction technique. We observed S11 reflection parameters of the sensor at resonant frequencies depend on the glucose concentration. We could determine the glucose concentration in the 0-250 mg/dl concentration range at an operating frequency near 2.6 GHz. The measured minimum detectable signal was 0.018 dB/(mg/dl) and the measured minimum detectable concentration was 1.68 mg/dl. The simulation result for the minimum detectable signal and the minimum detectable concentration was 0.023 dB/(mg/dl) and 1.3 mg/dl, respectively. Our proposed system has excellent potential to serve as a human bloodless glucometer.
oai:arar.sci.am:23542
ՀՀ ԳԱԱ Հիմնարար գիտական գրադարան
Dec 13, 2023
Feb 27, 2020
44
https://arar.sci.am/publication/26310
Edition name | Date |
---|---|
Glucose Aqueous Solution Sensing by Modified Hilbert Shaped Microwave Sensor | Dec 13, 2023 |
A. O. Vardanyan S. R. Melikyan K. A. Movsesyan D. L. Oganesyan E. V. Chakhoyan
H. D. Khondkaryan
Zh. Baghdasaryan Babajanyan, A. L. Odabashyan H. Lee B. Friedman K. Lee
Conference Chairs: Arsen Hakhoumian, (Dr.Sc., Ass. Member NAS RA, IRPhE, Armenia) Tigran Zakaryan (Dr. CEO IRPhE, Armenia)